首页 | 本学科首页   官方微博 | 高级检索  
     

原子荧光光谱法测定尿中微量砷元素
引用本文:杨志羡. 原子荧光光谱法测定尿中微量砷元素[J]. 化学分析计量, 2016, 0(5): 91-94. DOI: 10.3969/j.issn.1008-6145.2016.05.024
作者姓名:杨志羡
作者单位:内蒙古地方病防治研究中心,呼和浩特,100031
摘    要:
建立了测定尿中微量砷元素含量的原子荧光光谱法。以硝酸–高氯酸(4∶1)混合液消解样品,加入5%的硫脲–抗坏血酸混合液还原样品消解液,将处理好的样品导入原子荧光仪进行测定,荧光强度I_f与砷元素质量浓度c线性相关,线性方程为I_f=178.351c+3.131,相关系数为0.999 8,砷的检出限为0.024 3μg/L,测定结果的相对标准偏差为3.0%~4.1%(n=9),加标回收率为100%~104%。该法样品处理简便,适于大批量尿样中砷含量的检测。

关 键 词:原子荧光光谱  尿样    样品前处理

Determination of Trace Arsenic in Urine by Atomic Fluorescence Spectrometry
Abstract:
An atomic fluorescence spectrometry for determining arsenic content in urine was set up. Nitric acid–perchloric acid (4∶1) was used to digest the urine sample, and the sample solution was reducted by adding 5% thiourea–ascorbic acid solution, the digested samplewas determined by atomic fluorescence spectrometer. The fluorescence enhencity was linear with the concentration of arsenic, the linear equation wasIf=178.351c+3.131 with the corelatopn coefficient of 0.999 8. The detection limit of arsenic was 0.024 3μg/L, the relative standard deviation of determination results were 3.0%–4.1%(n=9), and the recoveries were 100%–104%. This method has the advantage of simple pretreatment method for sample, and it was suitable for determining large batch of urine samples.
Keywords:atomic fluorescence spectroscopy  urine samples  arsenic  sample pretreatment
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号