Determination of uranium by adsorptive stripping voltammetry at a lead film electrode |
| |
Authors: | Mieczyslaw Korolczuk Katarzyna Tyszczuk Malgorzata Grabarczyk |
| |
Affiliation: | Faculty of Chemistry, Maria Curie-Sklodowska University, 20-031 Lublin, Poland |
| |
Abstract: | An adsorptive stripping voltammetric procedure for the determination of U(VI) at an in situ plated lead film electrode is described. The U(VI) complex with cupferron was accumulated from an acetate buffer solution of pH 4.2 at the potential −0.65 V. The measurements were carried out from undeaerated solutions. The calibration graph for an accumulation time of 180 s was linear from 5 × 10−10 to 2 × 10−8 mol L−1. The detection limit was 2 × 10−10 mol L−1, the relative standard deviation for 2 × 10−8 mol L−1 U(VI) was 4.3%. The proposed procedure was validated in the course of U(VI) determination in water certified reference materials. |
| |
Keywords: | Lead film electrode Stripping analysis Uranium Determination |
本文献已被 ScienceDirect 等数据库收录! |