首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A novel Mn PVC membrane electrode based on a recently synthesized Schiff base
Authors:Mohammad Hossein Mashhadizadeh  Ehsan Pour Taheri
Institution:a Faculty of Chemistry, Tarbiat Moallem University, Tehran, Iran
b Chemistry Department, Shahid Bahonar University of Kerman, Kerman, Iran
Abstract:A new PVC membrane electrode for manganese(II) ion based on a recently synthesized Schiff base of 5-(4-nitrophenylazo)-N-hexylamine]salicylaldimine is reported. The electrode exhibits a Nernstian response for Mn2+ ions over a wide concentration range (4.0 × 10−7 to 1.8 × 10−2 mol L−1) with a slope of 30.1 (±1.0). The limit of detection is 1.0 × 10−7 mol L−1. The electrode has a fast response time (∼10 s), a satisfactory reproducibility and relatively long life time. The proposed sensor revealed good selectivities over a wide variety of other cations include hard and soft metals. This electrode could be used in a pH range of 4.5-7.5. It was used as an indicator electrode in potentiometric titration of manganese(II) ions with EDTA solution.
Keywords:Manganese(II)-selective electrode  Schiff base  PVC membrane electrode  Potentiometry
本文献已被 ScienceDirect 等数据库收录!