Abstract: | A new class of monocrystalline diamond paste-based electrodes is proposed for the determination of chromium(III) at trace levels in vitamins. Three types of monocrystalline diamond—natural diamond 1 (natural diamond), synthetic diamond 50 (synthetic-1), and synthetic diamond 1 (synthetic-2)—were used for electrode construction. The linear concentration ranges are between 10–10 and 10–8; 10–9 and 10–7, and 10–10 to 10–8 mol L–1, with limits of detection of 10–12, 10–12, and 10–11 mol L–1, when natural diamond, synthetic-1, and synthetic-2, respectively, are used as electrode materials. For electrodes based on natural diamond and synthetic-1 it was found that Cr(III) yields a peak at about +0.275±0.015 V (vs. Ag/AgCl) within a predetermined positive potential range situated between +0.4 and +0.2 V, while for the electrode based on synthetic-2 the peaks are found at +0.300±0.015 V (vs. Ag/AgCl). The proposed method is reliable for the determination of chromium(III) at trace levels in two vitamin tablets (RSD<0.2%). |