Columns and optimum gradient conditions for fast second-dimension separations in comprehensive two-dimensional liquid chromatography
Authors:Tomá? Hájek  Pavel Jandera
Institution:Department of Analytical Chemistry, University of Pardubice, Pardubice, Czech Republic.
Abstract:Gradient elution provides significantly higher peak capacity in comparison to the isocratic elution mode, hence it is very useful in online comprehensive two-dimensional liquid chromatography (LC). We compared suitability of five commercial core-shell columns and one monolithic column for fast gradients in the second LC dimension, where the time of separation is strictly limited by the fraction cycle time. In two-dimensional reversed-phase systems with partially correlated retention, the resolution, the peak capacity, and the regularity of coverage of the second-dimension retention space can be improved by appropriate adjusting the gradient time and the gradient range to suit the sample properties. We developed a new strategy for adjusting the gradient mobile phase composition range in the second-dimension, employing the retention data of representative sample standards characterizing the sample properties, which can be calibrated using the reference alkylbenzene series. Optimized second-dimension gradients with single-step or segmented profiles covering two or more fraction ranges, employed for the separation of subsequent fractions from the first-dimension, improve significantly the resolution, the separation time, and the regularity of coverage of the two-dimensional retention plane. The approach was applied to the two-dimensional comprehensive separation of phenolic acids and flavonoid compounds occurring as natural antioxidants.
Keywords:Core‐shell columns  Fast second‐dimension gradients  Flavones  Phenolic acids  Two‐dimensional comprehensive liquid chromatography
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号