Detection efficiency calculations using Geant4 for a broad-energy germanium gamma spectrometer |
| |
Authors: | Chuanlei Liu Kurt Ungar Dylan Pierce Ian Hoffman Weihua Zhang |
| |
Affiliation: | 1.Radiation Protection Bureau of Health Canada,Ottawa,Canada;2.Royal Military College of Canada,Kingston,Canada |
| |
Abstract: | In assistance of radionuclide measurements at Canada’s Comprehensive Nuclear-Test-Ban Treaty (CTBT) laboratory, a Geant4 Monte Carlo application has been developed in simulating a broad-energy germanium detector and calculating detection efficiencies. The detector model was optimized in a reliable and non-biased manner through simultaneous tuning on gap distance and detector dimension, and was validated over various realistic measurement scenarios. All work is based on a series of experiments which covers the typical energy range of gamma radiation in environmental analysis, and considers the variety of the CTBT sample type, dimension and distance-to-detector. In all cases, the predicted efficiencies are consistent with the empirical ones within 5%, with a typical deviation of 3% in majority. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|