首页 | 本学科首页   官方微博 | 高级检索  
     


Enzymatic detection of glucose using fluoride-selective electrodes with polymeric membranes
Authors:Łukasz Górski  Dorota Klimaszewska  Mariusz Pietrzak  Elżbieta Malinowska
Affiliation:(1) Department of Analytical Chemistry, Faculty of Chemistry, Warsaw University of Technology, ul. Noakowskiego 3, Warsaw, 00-664, Poland
Abstract:The feasibility of using polymeric membrane fluoride-selective electrodes based on zirconium(IV) 5,10,15,20-tetraphenylporphyrin as a detector in a flow-injection analysis (FIA) system for glucose determination was examined. The optimization of enzymatic reactions, FIA system configuration and enzyme-immobilization process was performed. It was shown that the resulting flow-injection system exhibits good working parameters, such as reproducibility, linear range of glucose concentration (3 × 10−3–10−1 M), sampling rate (60 samples per minute) and lifetime (over 1 month). The performance of the polymeric membrane electrode was similar to that of a crystalline LaF3 electrode. The results of glucose determination in synthetic samples with the proposed system show good agreement with real glucose concentrations.
Keywords:Ion-selective electrodes  Biosensors  Chemical sensors  Enzymes  Flow injection