首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of gatifloxacin in human plasma by liquid chromatography/electrospray tandem mass spectrometry
Authors:Vishwanathan K  Bartlett M G  Stewart J T
Affiliation:Department of Pharmaceutical and Biomedical Sciences, College of Pharmacy, The University of Georgia, Athens, GA 30602-2352, USA.
Abstract:Gatifloxacin is an advanced-generation, 8-methoxyfluoroquinolone that is active against a broad spectrum of pathogens, including antiobiotic resistant Streptococcus pneumoniae. Development of a rapid, sensitive and selective method for the determination of gatifloxacin in human plasma is essential for understanding the pharmacokinetics of the drug when administered orally or intravenously. Solid phase extraction (SPE) using Oasis HLB was used to extract gatifloxacin and the internal standard ciprofloxacin from plasma. A method based on liquid chromatography/electrospray tandem mass spectrometry (LC/ESI-MS/MS) was developed and validated to quantitate gatifloxacin in human plasma. The precursor and major product ions of the analyte were monitored on a triple quadrupole mass spectrometer with positive ion electrospray ionization (ESI) in the multiple reaction monitoring (MRM) mode. Mechanisms for the formation of collision-induced dissociation products of gatifloxacin are proposed. Linear calibration curves were generated from 10--1000 ng/mL with coefficients of determination greater than 0.99. The interday and intraday precision (%RSD) was less than 6.0% and accuracy (%error) was less than 5.4% for gatifloxacin. The limit of detection (LOD) for the method was 500 pg/mL based on a signal-to-noise ratio of 3.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号