Potentiometric flow-injection determination of boron by using a flow-through tetrafluoroborate ion-selective poly(vinyl chloride) membrane electrode
Authors:Toshihiko Imato  Takeshi Yoshizuka  Nobuhiko Ishibashi
Affiliation:Department of Applied Analytical Chemistry, Faculty of Engineering, Kyushu University, Hakozaki, Higashi-ku, Fukuoka 812 Japan
Abstract:A highly sensitive tetrafluoroborate ion-selective poly(vinyl chloride) membrane electrode was constructed and applied to a detector in the flow-injection determination of boron. Boron at the 10 ng ml?1 level can be determined by converting it to tetrafluoroborate ion in a flow of hydrofluoric acid. A sampling rate of 30 h?1 was achieved with a relative standard deviation of less than 2%.