Evaluation of the dosimetric characteristics of a radiophotoluminescent glass dosimeter for high-energy photon and electron beams in the field of radiotherapy
Authors:Kihong Son  Haijo Jung  Sang Hun Shin  Hyun-Ho Lee  Mi-Sook Kim  Young Hoon Ji  Kum Bae Kim
Institution:aDepartment of Radiological Cancer Medicine, University of Science and Technology, Republic of Korea;bDivision of Radiation Cancer Research, Research Institute of Radiological and Medical Sciences, Republic of Korea;cDepartment of Radiation Oncology, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences, Seoul, Republic of Korea
Abstract:We aimed to evaluate the suitability of a glass dosimeter (GD) for high-energy photon and electron beams in experimental and clinical use, especially for radiation therapy. We examined the expanded dosimetric characteristics of GDs including dose linearity up to 500 Gy, uniformity among GD lots and for individual GDs, the angular dependence, and energy dependence of 4 therapeutic x-ray qualities. In addition, we measured the dosimetric features (dose linearity, uniformity, angular dependence, and energy dependence) of the GD for electron beams of 10 different electron energy qualities. All measurements with the exception of dose linearity for photon beam were performed in a water phantom. For high-energy photon beams, dose linearity has a linear relationship for a dose ranging from 1 to 500 Gy with the coefficient of determination; R2 of 0.998. The uniformity of each GD of dose measurements was within ±0.5% for four GD lots and within ±1.2% for 80 GDs. In terms of the effects of photon beam angle, lower absorbed doses of within 1.0% were observed between 60° and 105° than at 90°. The GD energy dependence of 4 photon beam energy qualities was within ±2.0%. On the other hand, the result of the dose linearity for high-energy electron beams showed well fitted regression line with the coefficient of determination; R2 of 0.999 between 6 and 20 MeV. The uniformity of GDs exposed to the nominal electron energies 6, 9, 12, 16, and 20 MeV was ±1.2%. In terms of the angular dependence to electron beams, absorbed doses were within 2.0% between 60° and 105° than at 90°. In evaluation of the energy dependence of the GD at nominal electron energies between 5 and 20 MeV, we obtained responses between 1.1% and 3.5% lower than that for a cobalt-60 beam. Our results show that GDs can be used as a detector for determining doses when a high-energy photon beam is used, and that it also has considerable potential for dose measurement of high-energy electron beam.
Keywords:Glass dosimeter  Dosimetric characteristics  Photon beam  Electron beam
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号