Ion chromatographic determination of iodide in urine and serum using a tubular ion-selective electrode based on a homogeneous crystalline membrane |
| |
Authors: | Agostinho A. Almeida Xiao Jun José L. F. C. Lima |
| |
Affiliation: | (1) CEQUP/Departamento de Química-Física, Faculdade de Farmácia, Universidade do Porto, Rua Aníbal Cunha, 164, 4050 Porto, Portugal |
| |
Abstract: | The use of a laboratory-made iodide ion-selective electrode with tubular configuration and based on a crystalline membrane (AgI/Ag2S) as the detector for ion chromatographic determination of iodide in urine and serum is described. A CIS reversed-phase column was coated withN-cetylpyridinium chloride to prepare a low-exchange-capacity analytical column and with hexadecyltrimethylammonium bromide to prepare a concentrator pre-column. A 2.0 ml min–1 flow rate of deionized water and 0.1 mol 1–1 KNO3 solution was used for the pre-concentration and for the chromatographic separation, respectively. For optimum performance of the detector a background level of iodide was added into the column effluent. A linear relationship (r = 0.9997) between tubular electrode potential (as peak height) and iodide concentration in the range 5–400 g 1–1 and a detection limit of 1.47 g 1–1 were obtained. The method shows good reproducibility for both peak height (2.2% RSD) and retention time (1.3% RSD). Recoveries on its application to the samples were 93.0–100.9% for urine and 91.4–106.0% for serum. |
| |
Keywords: | iodide ion chromatography tubular potentiometric detector urine serum |