Solid State Analysis with the New Leipzig High-Energy Ion Nanoprobe
Authors:Jürgen Vogt  Rolf-Horst Flagmeyer  Johannes Heitmann  Dietmar Lehmann  Tilo Reinert  Steffen Jankuhn  Daniel Spemann  Wolfgang Tröger  Tilman Butz
Institution:Universit?t Leipzig, Fakult?t für Physik und Geowissensch aften, Abt. Nukleare Festk?rperphysik, Linnéstr. 5, D-04103 Leipzig,
Abstract: The high-energy ion nanoprobe LIPSION at the University of Leipzig has been operational since October 1998. The ultrastable single ended 3.5 MV SINLETRONTM accelerator supplies the H+ or He+ ion beam. A magnetic scanning system moves the focused beam across the sample. At present, a resolution of 150 nm in the low current mode and 300 nm at 5 pA could be achieved. The UHV grade experimental chamber is equipped with electron-, energy dispersive X-ray-, and particle detectors. They can be used simultaneously to analyse the sample by means of PIXE (particle induced X-ray emission), RBS (Rutherford backscattering) and in the case of thin samples STIM (scanning transmission ion microscopy). A goniometer allows the application of channeling measurements in single crystals in combination with these methods. The detection limits depend on the elements to be analysed and range from (1000⋯1) μg/g relative and (1⋯0.01) pg absolute. The analysis is nondestructive, but the sample has to be vacuum resistant. Applications of the nanoprobe in the field of semiconductor research, biomedicine, and archaeology will be described.
Keywords::   Ion beam analysis  ion microprobe  solid state analysis  microanalysis  