Determination of cadmium in tobacco smoke and zinc in tap water by solvent extraction flame atomic absorption spectrometry
Authors:Francisco García Sánchez  Aurora Navas Díaz  Angel Arbáizar
Institution:(1) Departamento de Quimica Analitica, Facultad de Ciencias, Universidad de Málaga, E-29071 Málaga, Spain
Abstract:Procedures are described for the determination of cadmium in tobacco smoke and zinc in tap water by flame atomic absorption spectrometry after chelation with benzyl 2-pyridyl ketone 2-pyridylhydrazone and extraction with isobutyl methyl ketone. Calibration graphs in the organic layer were linear up to 1.3 mgrg/ml, with sensitivities of 80 ng/ml (Zn) and 71ng/ml (Cd), and limits of detection of 49 ng/ml for zinc and 38 ng/ml for cadmium. RSD values were 2.9% at 0.42 mgrg/ml (Cd) and 4.7% at 0.17 mgrg/ml (Zn).
Keywords:cadmium  zinc  atomic absorption spectrometry  tobacco smoke  tap water  BPKPH extraction