Determination of metal-ligand stoichiometries for inorganic complexes using total reflection X-ray fluorescence |
| |
Authors: | Greaves E D Bennun L Gomez J Nemeth P Sajo-Bohus L |
| |
Affiliation: | Universidad Simón Bolívar, Sección de Física Nuclear, Apartado 89000, Caracas, Venezuela. egreaves@usb.ve |
| |
Abstract: | The methods usually used to determine the ratio metal-ligand in inorganic complexes require a set of solutions with different concentrations for both the ligand and metal. We propose a new method using the total reflection X-ray fluorescence technique, in which the ratio between metal and ligand is determined precisely, easily, and quickly. Experimental results provide evidence that for different chemical complexes, the ligand-metal ratio determined by this technique deviates at most from stoichiometric values by 6%. The technique is restricted usually to elements with Z above 14, and its detection limit is on the order of 10(-8) g/g. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|