首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation and validation of the ABI 3700, ABI 3100, and the MegaBACE 1000 capillary array electrophoresis instruments for use with short tandem repeat microsatellite typing in a forensic environment
Authors:Koumi Pieris  Green Helen E  Hartley Susan  Jordan Darren  Lahec Sharon  Livett Richard J  Tsang Kam W  Ward David M
Institution:The Forensic Science Service, Solihull, UK. pku@fss.org.uk
Abstract:The demand for high-throughput DNA profiling has increased with the introduction of national DNA databases and has led to the development of automated methods of short tandem repeat (STR) profile production; however, a potential bottleneck still exists at the gel electrophoresis stage. Capillary electrophoresis sequencers capable of processing 96 samples with considerably reduced manual intervention are now available. In this paper, we compare the ABI Prism 377 slab-gel sequencer currently used by the Forensic Science Service with three leading capillary array electrophoresis instruments: the ABI Prism 3700, the Amersham MegaBACE 1000 and the 16-capillary ABI Prism 3100. We describe the experiments used to evaluate and validate these platforms for forensic use with the STR multiplex Ampf/STR SGMplus 1, 2], along with comparative data from the ABI Prism 377 sequencer.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏