Analysis of testosterone in human urine using molecularly imprinted solid-phase extraction and corona discharge ion mobility spectrometry |
| |
Authors: | Mirmahdieh Shiva Mardihallaj Azam Hashemian Zahra Razavizadeh Jalal Ghaziaskar Hassan Khayamian Taghi |
| |
Affiliation: | Department of Chemistry, Isfahan University of Technology, Isfahan, Iran. |
| |
Abstract: | Analysis of testosterone was accomplished using corona discharge ion mobility spectrometry. Molecular imprinted polymer was used for the extraction and pre-concentration of testosterone. Analytical parameters including precision, dynamic range and detection limit were obtained. The linear dynamic range was from 10 to 250 ng/mL and the limit of detection was 0.9 ng/mL. The proposed method was used for analysis of testosterone in urine samples. A urine sample from a 3-year-old girl was used as the blank. The RSD was below 10%. The obtained results from the method were also compared with the standard method for analysis of testosterone using SPE-HPLC analysis. The results demonstrate the accuracy of the method. |
| |
Keywords: | Ion mobility spectrometry Molecular imprinted polymer Testosterone |