An on-line potentiometric sequential injection titration process analyser for the determination of acetic acid
Authors:van Staden J F  Mashamba Mulalo G  Stefan Raluca I
Institution:Department of Chemistry, University of Pretoria, South Africa. Koos.vanStaden@chem.up.ac.za
Abstract:An on-line potentiometric sequential injection titration process analyser for the determination of acetic acid is proposed. A solution of 0.1 mol L(-1) sodium chloride is used as carrier. Titration is achieved by aspirating acetic acid samples between two strong base-zone volumes into a holding coil and by channelling the stack of well-defined zones with flow reversal through a reaction coil to a potentiometric sensor where the peak widths were measured. A linear relationship between peak width and logarithm of the acid concentration was obtained in the range 1-9 g/100 mL. Vinegar samples were analysed without any sample pre-treatment. The method has a relative standard deviation of 0.4% with a sample frequency of 28 samples per hour. The results revealed good agreement between the proposed sequential injection and an automated batch titration method.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!