Evaluation of a carbon-rod atomizer for routine determination of trace metals by atomic-absorption spectroscopy applications to analysis of lubricating oil and crude oil |
| |
Authors: | Hall G Bratzel M P Chakrabarti C L |
| |
Affiliation: | Department of Chemistry, Carleton University, Ottawa, Ontario K1S 5B6, Canada. |
| |
Abstract: | A carbon-rod atomizer (CRA) fitted with a 'mini-Massmann' carbon rod was evaluated for routine analysis of petroleum and petroleum products for trace metal content by atomic-absorption spectroscopy. Aspects investigated included sensitivity, detection limit, effect of solvent type, and interferences. The results of analysis of oil samples with this technique were compared with those obtained by other techniques. Metals studied were silver, copper, iron, nickel, and lead. Sensitivity and detection limit values obtained with the CRA were similar to those obtained with the carbon-filament atomizer. Strong 'solvent effects' were observed as well as interference by cations. On the basis of this study, design changes for the CRA are suggested, with the object of minimizing 'solvent effects' and interferences, increasing the atomization efficiency, and increasing the residence time of the atomic vapour in the optical path of the instrumental system. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|