Determination of bacterial antigens using a multichannel immunoenzyme amperometric sensor
Authors:G R Safina  E P Medyantseva  O N Bazarnova  N I Glushko  G K Budnikov
Institution:(1) Kazan State University, ul. Kremlevskaya 18, Kazan, 420008, Tatarstan, Russia;(2) Kazan Research Institute of Epidemiology and Microbiology, ul. K. Marksa 67, Kazan, 420015, Tatarstan, Russia
Abstract:A procedure for the individual determination of the antigens of the conditionally pathogenic micro-organisms Streptococcus pyogenes and Staphylococcus aureus present in combination using a multichannel immunoenzyme amperometric sensor based on a screen-printed electrode was developed. Conditions for fabricating the biosensing part and for the operation of the sensor (the matrix component, the ratio between biocomponents, and the substrate concentration) were chosen. The determination limits of bacterial antigens obtained using the analytical device were equal to 5 × 10?9 mg/mL. It was found that bacterial antigens simultaneously present in serums from patients can be determined using the sensor developed.
Keywords:
本文献已被 SpringerLink 等数据库收录!