首页 | 本学科首页   官方微博 | 高级检索  
     


On-line preconcentration of arsenic from water samples with an anion-exchanger column coupled to a hydride generation ICP system
Authors:Peter Schramel and Li-quiang Xu
Affiliation:(1) GSF Research Centre of Environment and Health, Institute of Ecological Chemistry, W-8042 Neuherberg, Federal Republic of Germany
Abstract:
Summary An on-line anion-exchange preconcentration hydride generation ICP system for the determination of total inorganic arsenic in water is described. The column was packed with strongly basic anion-exchange resin (AG 1-X8). Experimental conditions including pH of the sample solution, eluent, flow rate of eluent, oxidation states of arsenic and competing anion ions were studied. Compared with the conventional continuous hydride generation ICP, a 9.2-fold improvement in sensitivity was obtained with RSD 1–2% at 100 ng/ml. The detection limit (3sgr) was 0.08 ng/ml. The recoveries in water samples were satisfactory. The system provides complete automation of sample loading, eluting and regenerating of the resin.On leave from Shanghai Institute of Metallurgy, Chinese Academy of Sciences, Shanghai, China 200050
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号