首页 | 本学科首页   官方微博 | 高级检索  
     

红外热不对称分析用于面神经功能损伤的客观评估
引用本文:Liu XL,Hong WX,Song JL,Wu ZY. 红外热不对称分析用于面神经功能损伤的客观评估[J]. 光谱学与光谱分析, 2012, 32(3): 647-650
作者姓名:Liu XL  Hong WX  Song JL  Wu ZY
作者单位:燕山大学电气工程学院生物医学工程系;秦皇岛市中医医院
基金项目:中医药行业科研专项(200707022);国家自然科学基金项目(61074130)资助
摘    要:
健康人体的温度分布存在一种双侧对称性,当面神经功能受损时,将改变这种热分布的对称性。人体热量大多通过红外辐射的形式散发,因此利用红外热成像技术可以有效捕捉这种红外热辐射分布的改变。本文提出一种新的热不对称分析的方法——有效热面积比,即计算面部特定区域与对侧区域的温差乘以温度异常区面积与总面积之比。利用这种方法对面神经功能损伤的患者和健康人群作对照性试验,结果显示:此方法的特异性和敏感性平均为0.90和0.87,比常规方法提高了7%和26%,有效热面积比与面神经功能的损伤程度正相关,平均为0.664。因此,关于诊断和评估面神经功能,红外热成像是一种有力的工具,有效热面积比是一项高效的临床指标。

关 键 词:面神经功能  红外热不对称  红外热成像  评估

Using infrared thermal asymmetry analysis for objective assessment of the lesion of facial nerve function
Liu Xu-long,Hong Wen-xue,Song Jia-lin,Wu Zhen-ying. Using infrared thermal asymmetry analysis for objective assessment of the lesion of facial nerve function[J]. Spectroscopy and Spectral Analysis, 2012, 32(3): 647-650
Authors:Liu Xu-long  Hong Wen-xue  Song Jia-lin  Wu Zhen-ying
Affiliation:Department of Biomedical Engineering, College of Electrical Engineering, Yanshan University, Qinhuangdao 066004, China. liuxulong@ysu.edu.cn
Abstract:
The skin temperature distribution of a healthy human body exhibits a contralateral symmetry. Some lesions of facial nerve function are associated with an alteration of the thermal distribution of the human body. Since the dissipation of heat through the skin occurs for the most part in the form of infrared radiation, infrared thermography is the method of choice to capture the alteration of the infrared thermal distribution. This paper presents a new method of analysis of the thermal asymmetry named effective thermal area ratio, which is a product of two variables. The first variable is mean temperature difference between the specific facial region and its contralateral region. The second variable is a ratio, which is equal to the area of the abnormal region divided by the total area. Using this new method, we performed a controlled trial to assess the facial nerve function of the healthy subjects and the patients with Bell's palsy respectively. The results show: that the mean specificity and sensitivity of this method are 0.90 and 0.87 respectively, improved by 7% and 26% compared with conventional methods. Spearman correlation coefficient between effective thermal area ratio and the degree of facial nerve function is an average of 0.664. Hence, concerning the diagnosis and assessment of facial nerve function, infrared thermography is a powerful tool; while the effective ther mal area ratio is an efficient clinical indicator.
Keywords:Facial nerve function  Infrared thermal asymmetry  Infrared thermography  Assessment
本文献已被 CNKI PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号