Enzymatic flow-injection determination of urea in blood serum using potentiometric gas sensor with internal nonactin based ISE
Authors:Krawczyński vel Krawczyk T  Trojanowicz M  Lewenstam A
Institution:Department of Chemistry, University of Warsaw, Pasteura 1, Warsaw, Poland.
Abstract:Ion-selective electrode with cellulose triacetate membrane containing nonactin is employed for the potentiometric detection of ammonia produced in biocatalytic reaction in flow-injection system with enzyme reactor. The elimination of interferences occurring in the presence of alkali metal ions was achieved by covering a nonactin membrane with outer hydrophobic gas permeable membrane. The obtained flow-injection response to ammonia indicates a possibility of ammonia determination down to 10 microM ammonia. In the flow-injection system for urea determination 200 microl of 10-fold diluted blood serum sample was injected into carrier stream of distilled water merged with TRIS buffer, passed through the urease flow-through reactor and then after merging with NaOH stream delivered to the detector. It was found in several series of natural blood serum samples, that the correction for endogenous ammonia in such a determination is not indispensable.