首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Continuous automatic determination of ammonia by using an integrated separation/detection unit
Authors:de la Torre M  de Castro M D  Valcarcel M
Institution:Department of Analytical Chemistry, Faculty of Sciences, University of Córdoba, 14004 Córdoba, Spain.
Abstract:An automatic continuous-flow photometric method for the determination of ammonia is proposed. It is based on the Berthelot reaction, the product of which is temporarily immobilized on a suitable support (Sephadex QAE) located in the flow-cell. The retained product is eluted after measurement by a cationic surfactant contained in the carrier solution. The method allows the determination of the analyte between 0.4 and 20.0 gmg/ml, with an RSD of 0.8%, at a sampling frequency of 13/hr. The determination limit can be reduced by a factor of ten by using a flow-cell with a two-fold longer path length, but the sampling frequency is also reduced as a result. The method was applied to the determination of this analyte in agricultural samples (plants and soils) and compared with standard methods for these types of samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号