Discontinuous-flow potentiometric determination of chloride in a large-volume wall-jet cell using an ion-selective electrode based on a silver chloride film chemically deposited on a silver iodide support |
| |
Authors: | Lexa J Stulík K |
| |
Affiliation: | Department of Chemistry and Ecology, Skoda Research, Tylova 46, 316 00 Pilsen, Czech Republic. |
| |
Abstract: | A robust and sensitive chloride ion-selective electrode can be prepared by modifying the surface of an iodide-selective electrode using the chemical reaction with mercuric chloride in an oxidizing medium containing excess chloride. A thin film of silver chloride is thus formed ensuring a rapid and reproducible response to chloride. The analytical parameters of this electrode are similar to those of commercial silver chloride ion-selective electrodes, but its electrical impedance and signal noise are substantially lower and the response somewhat faster. Its sensitivity toward surfactants is somewhat suppressed. The electrode was used for discontinuous flow potentiometric (DFP) determinations in a large-volume wall-jet cell in which the electrode surface can be continuously reactivated by a cleaning solution contained in the cell. The method was applied to determination of chloride in ground waters from an industrial waste dumping site. The limit of determination is low 9 mug Cl(-)/l (2.6 x 10(-7)M), the precision good (the relative standard deviation varies from 0.6 to 3.0% for chloride contents from 2.90 to 0.15 mg/l, respectively) and the method correlates satisfactorily with the results of an indirect AAS determination of chloride. The sample throughput is high-90 measurements can be carried out per hour, corresponding to 30-40 determinations per hour. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|