Comparison of four microwave digestion methods for the determination of selenium in fish tissue by using hydride generation atomic absorption spectrometry
Authors:Lan W G  Wong M K  Sin Y M
Affiliation:Department of Chemistry, National University of Singapore, 10 Kent Ridge Crescent, Singapore 0511, Republic of Singapore.
Abstract:Four microwave digestion methods of fish tissue for selenium determination by hydride generation atomic absorption spectrometry were compared, in which potassium hexacyanoferrate(III) was chosen as a masking agent for eliminating matrix interferences. The results showed that the methods employing HNO(3)/H(2)O(2), HNO(3)/K(2)S(2)O(8)/H(2)O(2) and HNO(3)/H(3)PO(4)/H(2)O(2) digestion media were unreliable. However, the decomposition using the digestion media of HNO(3)/H(2)SO(4)/H(2)O(2) enabled adequate digestion of fish tissue and retention of selenium in a state amenable for determination. Therefore, the digestion procedures with HNO(3)/H(2)SO(4)/H(2)O(2) media are proposed for the determination of selenium in fish tissue by hydride generation atomic absorption spectrometry. The recoveries of the spiked samples investigated ranged from 90 to 102%. The result obtained from analyzing the NIES CRM No. 6 mussel was in good agreement with the reference value (reference value: 1.5 mug/g; found: 1.45 +/- 0.05 mug/g). The limit of detection for selenium was 0.03 mug/g dry mass for a 100 mg sample. The contents of selenium in local fish species investigated ranged from 0.49 to 2.90 mug/g, and the relative standard deviation for the determination of selenium was less than 8%.
Keywords:
本文献已被 PubMed 等数据库收录!