Abstract: | ![]() An adsorptive stripping voltammetry method for the determination of traces of molybdenum(VI) in flowing solution at a wall-jet electrode sensor has been developed. After adsorption of a molybdenum complex on a wall-jet mercury film electrode, the complex is reduced by a square wave scan. More satisfactory results were obtained using 8-hydroxyquinoline as a complexing agent in nitrate medium than using Toluidine Blue in oxalic acid. Enhanced sensitivity was achieved by optimizing adsorption time and square wave parameter values. The detection limit of Mo(VI) was found to be at the nanomolar level. Interference of some other metallic species in the determination of nanomolar Mo(VI) was also investigated: Cu(II), Zn(II), Mn(II) do not interfere at 10 muM, whereas 1 muM FeEDTA(-) causes an increase in peak current. This iron interference was removed effectively with citric acid. |
| |