首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chemiluminescence flow system for the determination of sulfite
Authors:Wei Qin  Z-J Zhang and Cheng-Jie Zhang
Institution:(1) Department of Chemistry, Nanjing University, Nanjing 210008, P.R. China, CN;(2) Department of Chemistry, Shaanxi Normal University, Xi’an 710062, P.R. China, CN
Abstract:A novel chemiluminescence(CL) flow system for sulfite is described based on electrostatically immobilized luminol on an anion exchange column. Sulfite is detected by the CL reaction with luminol bleeding from the column by hydrolysis. The calibration graph is linear in the range 3 × 10–7 to 1 × 10–5 mol/L, and the detection limit is 1 × 10–7 mol/L. Interfering metal ions co-existing in sample solutions could be effectively eliminated on-line by an upstream cation exchanger. A complete analysis could be performed in 1 min with a relative standard deviation of less than 5%. The system could be reused for over 50 h and has been applied successfully to the determination of sulfur dioxide in air. Received: 21 October 1997 / Revised: 23 February 1998 / Accepted: 26 February 1998
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏