首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Flow-injection spectrofluorimetric determination of ethylenethiourea
Authors:Tomás Pérez-Ruiz  Carmen Martínez-Lozano  Virginia Tomás  Antonio Sanz  Jesús Martín
Institution:(1) Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, E-30071 Murcia, Spain, ES
Abstract:A flow-injection configuration is proposed for the fluorimetric determination of ethylenethiourea. The procedure is based on the inhibitory effect of ethylenethiourea on the oxidation of thiamine to thiochrome by mercury(II). A linear calibration graph was obtained between 0.1 and 2.0 μg mL–1, with a sampling rate of 40 samples per hour and a relative standard deviation of about 1.11%. The usefulness of the method was tested for the determination of ethylenethiourea residues in water, milk, potatoes, pear, grape and apple. Received: 26 January 1998 / Revised: 6 April 1998 / Accepted: 9 April 1998
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号