A 532 nm molecular iodine optical frequency standard based on modulation transfer spectroscopy
Institution:MOE Key Laboratory of Fundamental Physical Quantities Measurement, Hubei Key Laboratory of Gravitation and Quantum Physics, PGMF and School of Physics, Huazhong University of Science and Technology, Wuhan 430074, China
Abstract:We report construction of an iodine-stabilized laser frequency standard at 532 nm based on modulation transfer spectroscopy (MTS) technology with good reproducibility. A frequency stability of $2.5 \times {10}^{-14}$ at 1 s averaging time is achieved, and the frequency reproducibility has a relative uncertainty of ${3.5\times }{10}^{-13}$, demonstrating the great stability of our setup. The systematic uncertainty of the iodine-stabilized laser frequency standard is evaluated, especially the contribution of the residual amplitude modulation (RAM). The contribution of the RAM in MTS cannot be evaluated directly. To solve this problem, we theoretically deduce the MTS signal with RAM under large modulation depth, and prove that the non-symmetric shape of the MTS signal is directly related to the MTS effect. The non-symmetric shape factor $r$ can be calibrated with a frequency comb, and in real experiments, this $r$ value can be obtained by least-squares fitting of the MTS signal, from which we can infer the RAMinduced frequency shift. The full frequency uncertainty is evaluated to be 5.3 kHz (corresponding to a relative frequency uncertainty of ${9.4\times }{10}^{-12})$. The corrected transition frequency has a difference from the BIPM-recommended value of 2 kHz, which is within ${1}\sigma$ uncertainty, proving the validity of our evaluation.
Keywords:iodine-stabilized laser frequency standard  modulation transfer spectroscopy  residual amplitude modulation  
本文献已被 CNKI 等数据库收录!
点击此处可从《中国物理 B》浏览原始摘要信息
点击此处可从《中国物理 B》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号