Micellar medium for the analysis of complex mixtures of molybdenum and tungsten by derivative synchronous spectrofluorimetry in steels |
| |
Authors: | Blanco C C Campaña A M Barrero F A Ceba M R |
| |
Institution: | Department of Analytical Chemistry, Faculty of Sciences, University of Granada, 18071 Granada, Spain. |
| |
Abstract: | The cationic surfactant hexadecyltrimethylammonium bromide enhances the fluorescence intensity of both molybdenum and tungsten complexes obtained with Alizarin Red S. This characteristic has been used for a sensitive simultaneous spectrofluorimetric determination of both ions over the range 32-4000 ng/ml and 49-2500 ng/ml for Mo and W, respectively. Good selectivity is achieved applying the well known synchronous derivative technique with a minucious pH selection. The proposed method has been applied to a certified steel sample with recoveries between 96 and 102%. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|