首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Gas chromatography/mass spectrometry of polychlorinated biphenyls using atmospheric pressure chemical ionization and atmospheric pressure photoionization microchips
Authors:Luosujärvi Laura  Karikko Mika-Matti  Haapala Markus  Saarela Ville  Huhtala Sami  Franssila Sami  Kostiainen Risto  Kotiaho Tapio  Kauppila Tiina J
Institution:Laboratory of Analytical Chemistry, University of Helsinki, Finland.
Abstract:Gas chromatography (GC) and ion trap mass spectrometry (MS) were combined with microchip atmospheric pressure chemical ionization (microAPCI) and microchip atmospheric pressure photoionization (microAPPI) sources. Selected polychlorinated biphenyls (PCBs, IUPAC Nos. 28, 52, 101, 118, 138, 153 and 180) were analyzed by GC/microAPCI-MS and GC/microAPPI-MS to demonstrate the applicability of the miniaturized ion sources in negative ion mode analysis. The microAPCI and microAPPI methods were evaluated in respect of detection limit, linearity and repeatability. The detection limits for the PCB congeners were somewhat lower with microAPCI than with microAPPI, whereas microAPPI showed slightly wider linear range and better repeatability. With both methods, the best results were obtained for highly chlorinated or non-ortho-chlorinated PCBs, which possess the highest electron affinities. Finally, the suitability of the GC/microAPPI-MS method for the analysis of PCBs in environmental samples was demonstrated by analyzing soil extracts, and by comparing the results with those obtained by gas chromatography with electron capture detection (GC/ECD).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏