首页 | 本学科首页   官方微博 | 高级检索  
     


The limit of detection in differential scanning calorimetry
Authors:S. Wies  A. Geyer  W. Eysel
Affiliation:(1) Mineralogisch-Petrographisches Institut, Universität Heidelberg, F.R.G.
Abstract:A procedure is described to determine the limit of detection of DSC instruments by using tiny signals from spontaneous polymorphic transitions of CsCl, K2Cr2O7 and Na2SO4. It is shown how such signals can be found well-resolved in DSC diagrams of powder samples. To distinguish them from the baseline noise they should exhibit a height at least twice that of the baseline width. For the instrument employed the corresponding smallest amount of heat, i.e., the limit of detection, was found to be 0.1 mJ.The authors thank Mr. H. Maltry for technical help and the Deutsche Forschungsgemeinschaft for support.
Keywords:DSC  limit of detection  polymorphic transitions
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏