首页 | 本学科首页   官方微博 | 高级检索  
     


Optimization of conductivity detection of low-molecular-mass anions in capillary zone electrophoresis
Authors:Martin U. Katzmayr   Christian W. Klampfl  Wolfgang Buchberger
Affiliation:

Department of Analytical Chemistry, Johannes-Kepler-University Linz, Altenbergerstrasse 69, A-4040 Linz, Austria

Abstract:The optimization of background electrolyte compositions for capillary zone electrophoresis in combination with conductivity detection focusing on maximal detector response is discussed. A theoretical approach pointing out the influence of the electrolyte co- and counter-ion mobilities on the detector signal has been developed. Using this model, running buffer compositions providing optimum S/N ratios for the selected analytes could be calculated. The results derived from these examinations have been verified by experimental investigations, namely the determination of inorganic and organic anionic solutes.