首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Isotope selective analysis of CO(2) with tunable diode laser (TDL) spectroscopy in the NIR
Authors:Horner Gerald  Lau Steffen  Kantor Zoltan  Lohmannsroben Hans-Gerd
Institution:Institute of Chemistry, Physical Chemistry, Potsdam University, Karl-Liebknecht Str. 24-25, D-14476 Potsdam-Golm, Germany.
Abstract:The performance of a home-built tunable diode laser (TDL) spectrometer, aimed at multi-line detection of carbon dioxide, has been evaluated and optimized. In the regime of the (30(0)1)(III) <-- (000) band of (12)CO(2) around 1.6 microm, the dominating isotope species (12)CO(2), (13)CO(2), and (12)C(18)O(16)O were detected simultaneously without interference by water vapor. Detection limits in the range of few ppmv were obtained for each species utilizing wavelength modulation (WM) spectroscopy with balanced detection in a long-path absorption cell set-up. High sensitivity in conjunction with high precision -- typically +/-1 (per thousand) and +/-6 (per thousand) for 3% and 0.7% of CO(2), respectively -- renders this experimental approach a promising analytical concept for isotope-ratio determination of carbon dioxide in soil and breath gas. For a moderate (12)CO(2) line, the pressure dependence of the line profile was characterized in detail, to account for pressure effects on sensitive measurements.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号