首页 | 本学科首页   官方微博 | 高级检索  
     


Flow of the carrier gas in open-tubular capillary columns in temperature-programmed gas chromatography
Authors:Tibor Hevesi,J  n Krupcí  k
Affiliation:Tibor Hevesi,Ján Krupcík
Abstract:
A second-order non-linear partial different equation was derived to describe the dependence of carrier gas pressure in the column on the column distance and the time under temperature programmed conditions. This equation was solved numerically by the modified finite difference method for various column parameters. Constant inlet and outlet pressures were used as boundary conditions. The retention times calculated on assumption of a constant pressure profile along the column. Significant differences between retention times of corresponding solutes calculated by the two methods were found, especially when relatively long columns(L>50m) with small internal diameter(d<0.3mm)and high temperature program rates (r>5°/min)are used.
Keywords:Capillary gas chromatography  Temperature programmed gas chromatography  Modelling of separation  Carrier gas flow