首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of pethidine in human plasma by LC–MS/MS
Authors:Xianqin Wang  Zheng Xiang  Xiaojun Cai  Haiya Wu  Xuebao Wang  Junwei Li  Meiling Zhang
Institution:1. Analytical and Testing Center, Wenzhou Medical College, Wenzhou 325035, China;2. School of Pharmacy, Wenzhou Medical College, Wenzhou 325035, China;3. The Second Affiliated Hospital of Wenzhou Medical College, Wenzhou 325000, China
Abstract:A sensitive and selective liquid chromatography–tandem mass spectrometry method for the determination of pethidine in human plasma was developed and validated over the concentration range of 4–2000 ng/mL. After addition of ketamine as internal standard, liquid–liquid extraction was used to produce a protein‐free extract. Chromatographic separation was achieved on a 100 × 2.1 mm, 5 µm particle, AllureTM PFP propyl column, with 45:40:15 (v/v/v) acetonitrile–methanol–water containing 0.2% formic acid as mobile phase. The MS data acquisition was accomplished by multiple reactions monitoring mode with positive electrospray ionization interface. The lower limit of quantification was 4 ng/mL; for inter‐day and intra‐day tests, the precision (RSD) for the entire validation was less than 7%, and the accuracy was within 95.9–106.5%. The method is sensitive and simple, and was successfully applied to analysis of samples of clinical intoxication. Copyright © 2010 John Wiley & Sons, Ltd.
Keywords:pethidine  LC–  MS/MS  human plasma
设为首页 | 免责声明 | 关于勤云 | 加入收藏