A flow-based procedure with solenoid micro-pumps for the spectrophotometric determination of uric acid in urine |
| |
Authors: | Diogo L. Rocha |
| |
Affiliation: | Instituto de Química, Universidade de São Paulo, P.O. Box 26.077, 05599-970, São Paulo, SP, Brazil |
| |
Abstract: | The determination of uric acid in urine shows clinical importance, once it can be related to human organism dysfunctions, such as gout. An analytical procedure employing a multicommuted flow system was developed for the determination of uric acid in urine samples. Cu(II) ions are reduced by uric acid to Cu(I) that can be quantified by spectrophotometry in the presence of 2,2′-biquinoline 4,4′-dicarboxylic acid (BCA). The analytical response was linear between 10 and 100 μmol L− 1 uric acid with a detection limit of 3.0 μmol L− 1 (99.7% confidence level). Coefficient of variation of 1.2% and sampling rate of 150 determinations per hour were achieved. Per determination, 32 μg of CuSO4 and 200 μg of BCA were consumed, generating 2.0 mL of waste. Recoveries from 91 to 112% were estimated and the results for 7 urine samples agreed with those obtained by the commercially available enzymatic kit for determination of uric acid. The procedure required 100-fold dilution of urine samples, minimizing sample consumption and interfering effects. In order to avoid the manual dilution step, on-line sample dilution was achieved by a simple system reconfiguration attaining a sampling rate of 95 h− 1. |
| |
Keywords: | Multicommutation Flow analysis Clinical analysis Urine Uric acid |