Study of a pulsed glow discharge ion source for time-of-flight mass spectrometry |
| |
Institution: | 2. Max-Planck-Institut für Kernphysik, Heidelberg, Germany |
| |
Abstract: | This paper describes a new type of glow discharge (GD) ion source coupled to a time-of-flight mass spectrometer (TOFMS). The GD is operated in the microsecond pulse (μs-pulse) mode. The operational parameters of the μs-pulse GD were optimized against the ion signals, giving 180 Pa for the discharge pressure, 3 A for the transient discharge current, 1.75 kHz for the discharge frequency and 2 μs for the discharge pulse duration. Experimental results show that the discharge current in the μs-pulse mode can be one order of magnitude higher than that obtained in the d.c. mode. The structure of the interface between the μs-pulse GD and the mass spectrometer was found to be critical, and a Macor disc must be applied in front of the sampling orifice in order to shield the sampling plate from the anode of the GD to achieve both a good vacuum and the best sputtering. A transient sputtering rate of 24.4 μs s−1 mm−2 was reached in the μs-pulse mode and was significantly higher than that for the d.c.-GD. Typical mass spectra of brass and nickel samples were studied and are discussed. © 1997 Elsevier Science B.V. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|