Pyraflufen-ethyl residues in soil by solid phase extraction and high-performance liquid chromatography with UV detection. |
| |
Authors: | Hui-Li Wang Jian-Zhong Li Ji-Ye Hu |
| |
Affiliation: | Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, Beijing, China. |
| |
Abstract: | Solid-phase extraction (SPE) procedure for cleanup followed by HPLC-UV method has been investigated for the determination of pyraflufen-ethyl residues in soil. The pesticide is extracted from the sample with acetone-water (80:20, v/v) and the extract is loaded onto an octadecyl (C(18)) column. The pesticide is eluted with acetonitrile and determined by HPLC with a UV detector. Using an acetone-water extraction followed by a C(18) cleanup, this method is characterized by recovery >90.1%, precision <5.8% RSD and sensitivity of 0.01 mg/kg. The proposed method has been successfully employed for the determination of the degradation dynamics of pyraflufen-ethyl in four agricultural soil samples under laboratory conditions. |
| |