Dosimetry measurements at the fast neutron therapy facility in Seattle |
| |
Authors: | Ruedi Risler Alina Popescu |
| |
Affiliation: | University of Washington Medical Center, Seattle, WA 98195-6043, USA |
| |
Abstract: | The fast neutron therapy facility at the University of Washington has been in routine clinical use for 25 years. 50.5 MeV protons produce neutrons in a beryllium target mounted on an isocentric gantry. Beam shaping is accomplished with a 40-leaf collimator. Dosimetry measurements for treatment planning and calibration are performed with tissue equivalent ion chambers. A layered phantom of alternating Solid Water® and Plastic Water® slabs has been developed for rapid dose verification measurements. The neutron field in the room has been used for radiation testing of electronic components. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |