首页 | 本学科首页   官方微博 | 高级检索  
     


Design and validation of a single sphere multi-detector neutron spectrometer based on LiF: Mg,Cu,P thermoluminescent dosemeters
Authors:José María Gómez-Ros  Roberto Bedogni  Montserrat Moraleda  Ana Romero  Antonio Delgado  Adolfo Esposito
Affiliation:1. CIEMAT, Edif. 36, Av. Complutense 22, E-28040 Madrid, Spain;2. IFNF, LNF Frascati National Laboratory, U.F. Fisica Sanitaria, via E. Fermi n. 40, 00044 Frascati, Italy
Abstract:This communication describes a new neutron spectrometer consisting of pairs of 7Li and 6Li based thermoluminescent dosemeters (MCP-6, MCP-7) located at selected positions within a single moderating polyethylene sphere. The spatial arrangement of the dosemeters has been designed using the MCNPX Monte Carlo code to calculate the response matrix in order to obtain a nearly isotropic response for neutrons in the energy range up to 20 MeV. A partial validation of the calculated response matrix has been performed with the calibrated 241Am–Be neutron source at the INFN–LNF Laboratory, using the shadow cone technique.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号