Design and validation of a single sphere multi-detector neutron spectrometer based on LiF: Mg,Cu,P thermoluminescent dosemeters |
| |
Authors: | José María Gómez-Ros Roberto Bedogni Montserrat Moraleda Ana Romero Antonio Delgado Adolfo Esposito |
| |
Affiliation: | 1. CIEMAT, Edif. 36, Av. Complutense 22, E-28040 Madrid, Spain;2. IFNF, LNF Frascati National Laboratory, U.F. Fisica Sanitaria, via E. Fermi n. 40, 00044 Frascati, Italy |
| |
Abstract: | This communication describes a new neutron spectrometer consisting of pairs of 7Li and 6Li based thermoluminescent dosemeters (MCP-6, MCP-7) located at selected positions within a single moderating polyethylene sphere. The spatial arrangement of the dosemeters has been designed using the MCNPX Monte Carlo code to calculate the response matrix in order to obtain a nearly isotropic response for neutrons in the energy range up to 20 MeV. A partial validation of the calculated response matrix has been performed with the calibrated 241Am–Be neutron source at the INFN–LNF Laboratory, using the shadow cone technique. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|