首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simultaneous determination of vitamin B12 and its derivatives using some of multivariate calibration 1 (MVC1) techniques
Authors:Samadi-Maybodi Abdolraouf  Darzi S K Hassani Nejad
Institution:Analytical Division, Faculty of Chemistry, University of Mazandaran, Babolsar, Iran.
Abstract:Resolution of binary mixtures of vitamin B12, methylcobalamin and B12 coenzyme with minimum sample pre-treatment and without analyte separation has been successfully achieved by methods of partial least squares algorithm with one dependent variable (PLS1), orthogonal signal correction/partial least squares (OSC/PLS), principal component regression (PCR) and hybrid linear analysis (HLA). Data of analysis were obtained from UV-vis spectra. The UV-vis spectra of the vitamin B12, methylcobalamin and B12 coenzyme were recorded in the same spectral conditions. The method of central composite design was used in the ranges of 10-80mgL(-1) for vitamin B12 and methylcobalamin and 20-130mgL(-1) for B12 coenzyme. The models refinement procedure and validation were performed by cross-validation. The minimum root mean square error of prediction (RMSEP) was 2.26mgL(-1) for vitamin B12 with PLS1, 1.33mgL(-1) for methylcobalamin with OSC/PLS and 3.24mgL(-1) for B12 coenzyme with HLA techniques. Figures of merit such as selectivity, sensitivity, analytical sensitivity and LOD were determined for three compounds. The procedure was successfully applied to simultaneous determination of three compounds in synthetic mixtures and in a pharmaceutical formulation.
Keywords:UV–  vis spectrophotometry  PLS1  PCR  HLA  Orthogonal signal correction  Vitamin B12
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号