Joule heating in packed capillaries used in capillary electrochromatography
Authors:Rathore Anurag S  Reynolds Kimberly J  Colón Luis A
Institution:Technology Development, Bioprocess and Formulation Technology, Pharmacia Co., Chesterfield, MO 63198, USA. anurag.s.rathore@pharmacia.com
Abstract:Effective heat dissipation is critical for reproducible and efficient separations in electrically driven separation systems. Flow rate, retention kinetics, and analyte diffusion rates are some of the characteristics that are affected by variation in the temperature of the mobile phase inside the column. In this study, we examine the issue of Joule heating in packed capillary columns used in capillary electrochromatography (CEC). As almost all commonly used CEC packings are poor thermal conductors, it is assumed that the packing particles do not conduct heat and heat transfer is solely through the mobile phase flowing through the system. The electrical conductivity of various mobile phases was measured at different temperatures by a conductivity meter and the temperature coefficient for each mobile phase was calculated. This was followed by measurement of the electrical current at several applied voltages to calculate the conductivity of the solution within the column as a function of the applied voltage. An overall increase in the conductivity is attributed to Joule heating within the column, while a constant conductivity means good heat dissipation. A plot of conductivity versus applied voltage was used as the indicator of poor heat dissipation. Using theories that have been proposed earlier for modeling of Joule heating effects in capillary electrophoresis (CE), we estimated the temperature within CEC columns. Under mobile and stationary phase conditions typically used in CEC, heat dissipation was found to be not always efficient. Elevated temperatures within the columns in excess of 23 degrees C above ambient temperature were calculated for packed columns, and about 35 degrees C for an open column, under a given set of conditions. The results agree with recently published experimental findings with nuclear magnetic resonance (NMR) thermometry, and Raman spectroscopic measurements.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号