New calibration method for solid sampling Zeeman atomic absorption spectrometry (SS-ZAAS) for cadmium
Authors:C Hofmann  C Vandecasteele and J Pauwels
Institution:(1) Commission of the European Communities, Joint Research Centre, Central Bureau for Nuclear Measurements, Steenweg op Retie, B-2440 Geel, Belgium;(2) Department of Chemical Engineering, Katholieke Universiteit Leuven, de Croylaan 46, B-3001 Heverlee, Belgium
Abstract:Summary Calibration of solid sampling Zeeman AAS for cadmium was investigated. A series of certified reference materials was analysed using a CRM with a different composition as standard. For routine applications this yielded satisfactory results. A method of calibration was then applied involving addition of standard solution to lyophilized samples, followed by lyophilization and homogenization. Using an iteration procedure results were obtained which no longer depend on certified reference materials. The method was applied to the analysis of a testbatch of codfish powder prepared at CBNM. Finally, the spiked fish standards obtained in this way were also used to analyse other certified reference materials.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏