首页 | 本学科首页   官方微博 | 高级检索  
     


k0-NAA quality assessment by analysis of different certified reference materials using the KAYZERO/SOLCOI software
Authors:R. Jaćimović  B. Smodiš  T. Bučar  P. Stegnar
Affiliation:(1) Department of Environmental Sciences, Jo"zcaron"ef Stefan Institute, Jamova 39, P.O. Box 3000, SI-1000 Ljubljana, Slovenia;(2) Division of Human Health, International Atomic Energy Agency, Wagramer Strasse 5, P.O. Box 100, A-1400 Vienna, Austria
Abstract:A suite of natural matrix reference materials (RMs) were used to assess the quality of analytical results obtained by k0-instrumental neutron activation analysis (k0-INAA) at the Jozcaronef Stefan Institute (IJS). Five certified reference materials (CRMs) from the Institute for Reference Materials and Measurements (IRMM), two standard reference materials (SRMs) from the National Institute of Standards and Technology (NIST), three RMs from the International Atomic Energy Agency (IAEA) and one RM from IJS were analyzed. Altogether, results for twenty-four elements in inorganic matrices and twenty-nine elements in organic matrices, obtained by k0-INAA, were compared to certified values. Results obtained show good agreement with certified or assigned values except for Fe, La, Nd, Sm and U in inorganic matrices, and Ag, Al and Cr in organic matrices.
Keywords:
本文献已被 SpringerLink 等数据库收录!