1.
Recently, linear ion traps (LITs) have been combined with quadrupole (Q), time-of-flight (TOF) and Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometry (MS). LITs can be used either as ion accumulation devices or as commercially available, stand-alone mass spectrometers with MSn capabilities. The combination of triple quadrupole MS with LIT technology in the form of an instrument of configuration QqLIT, using axial ejection, is particularly interesting, because this instrument retains the classical triple quadrupole scan functions such as selected reaction monitoring (SRM), product ion (PI), neutral loss (NL) and precursor ion (PC) while also providing access to sensitive ion trap experiments. For small molecules, quantitative and qualitative analysis can be performed using the same instrument. In addition, for peptide analysis, the enhanced multiply charged (EMC) scan allows an increase in selectivity, while the time-delayed fragmentation (TDF) scan provides additional structural information. Various methods of operating the hybrid instrument are described for the case of the commercial Q TRAP (AB/MDS Sciex) and applications to drug metabolism analysis, quantitative confirmatory analysis, peptides analysis and automated nanoelectrospray (ESI-chip-MS) analysis are discussed.  相似文献   

2.
The quantitative capabilities of a linear ion trap high‐resolution mass spectrometer (LTQ‐Orbitrap™) were investigated using full scan mode bracketing the m/z range of the ions of interest and utilizing a mass resolution (mass/FWHM) of 15000. Extracted ion chromatograms using a mass window of ±5–10 mmu centering on the theoretical m/z of each analyte were generated and used for quantitation. The quantitative performance of the LTQ‐Orbitrap™ was compared with that of a triple quadrupole (API 4000) operating using selected reaction monitoring (SRM) detection. Comparable assay precision, accuracy, linearity and sensitivity were observed for both approaches. The concentrations of actual study samples from 15 Merck drug candidates reported by the two methods were statistically equivalent. Unlike SRM being a tandem mass spectrometric (MS/MS)‐based detection method, a high resolution mass spectrometer operated in full scan does not need MS/MS optimization. This approach not only provides quantitative results for compounds of interest, but also will afford data on other analytes present in the sample. An example of the identification of a major circulating metabolite for a preclinical development study is demonstrated. Copyright © 2009 John Wiley & Sons, Ltd.  相似文献   

3.
Bioanalytical support of plasma pharmacokinetic (PK) studies for drug discovery programs primarily involves the quantitative analysis of dosed compounds using liquid chromatography/atmospheric pressure ionization tandem mass spectrometry (LC/MS/MS) operated in selected reaction monitoring (SRM) mode. However, there is a growing need for information on the metabolism of new chemical entities (NCEs), in addition to the time-concentration profiles from these studies. In this paper, we present a novel approach to not only quantify parent drugs with SRM, but also simultaneously screen for metabolites using a hybrid triple quadrupole/linear ion trap (QqQ(LIT)) instrument. This was achieved by incorporating both the conventional SRM-only acquisition of parent compounds and the SRM-triggered information-dependent acquisition (IDA) of potential metabolites within the same scan cycle during the same LC/MS/MS run. Two test compounds were used to demonstrate the applicability of this approach. Plasma samples from PK studies were processed by simple protein precipitation and the supernatant was diluted with water before injection. The fast scanning capability of the linear ion trap allowed for the information-dependent acquisition of metabolite MS/MS spectra (<1 s/scan), in addition to the collection of adequate data points for SRM-only channels. The MS/MS spectra obtained from potential metabolites in post-dose samples correlated well with the spectra of the parent compounds studied, therefore providing additional confirmatory structure information without the need for repetitive analyses. Relative quantitative time-concentration profiles of identified metabolites were also obtained. Furthermore, this articulated SRM+SRM-IDA approach generated equivalent quantitative results for parent compounds to those obtained by conventional SRM-only analysis. This approach has been successfully used to support discovery PK screening programs.  相似文献   

4.
This paper describes a new strategy that utilizes the fast trap mode scan of the hybrid triple quadrupole linear ion trap (QqQ(LIT)) for the identification of drug metabolites. The strategy uses information-dependent acquisition (IDA) where the enhanced mass scan (EMS), the trap mode full scan, was used as the survey scan to trigger multiple dependent enhanced product ion scans (EPI), the trap mode product ion scans. The single data file collected with this approach not only includes full scan data (the survey), but also product ion spectra rich in structural information. By extracting characteristic product ions from the dependent EPI chromatograms, we can provide nearly complete information for in vitro metabolites that otherwise would have to be obtained by multiple precursor ion scan (prec) and constant neutral loss (NL) analysis. This approach effectively overcomes the disadvantages of traditional prec and NL scans, namely the slow quadrupole scan speed, and possible mass shift. Using nefazodone (NEF) as the model compound, we demonstrated the effectiveness of this strategy by identifying 22 phase I metabolites in a single liquid chromatography/tandem mass spectrometry (LC/MS/MS) run. In addition to the metabolites reported previously in the literature, seven new metabolites were identified and their chemical structures are proposed. The oxidative dechlorination biotransformation was also discovered which was not reported in previous literature for NEF. The strategy was further evaluated and worked well for the fast discovery setting when a ballistic gradient elution was used, as well as for a simulated in vivo setting when the incubated sample (phase I metabolites) was spiked to control human plasma extract and control human urine.  相似文献   

5.
The use of a new hybrid quadrupole/linear ion trap known as the Q TRAP offers unique benefits as a LC-MS-MS detector for both small and large molecule analyses. The instrument combines the capabilities of a triple quadrupole mass spectrometer and ion trap technology on a single platform. Product ion scans are conducted in a hybrid fashion with the fragmentation step accomplished via acceleration into the collision cell followed by trapping and mass analysis in the Q3 linear ion trap. This results in triple quadrupole fragmentation patterns with no inherent low molecular mass cutoff. In-trap fragmentation is also possible in order to provide triple MS (MS3) capabilities. There are also several scan modes that are not possible on conventional instruments that enable identification of analytes within complex biological matrixes for subsequent high sensitivity product ion scans. This report will describe the new hybrid instrument and the principles of operation, and also provide examples of the unique scan modes and capabilities of the Q TRAP for LC-MS-MS detection in metabolism identification.  相似文献   

6.
The application of a new hybrid RF/DC quadrupole-linear ion trap mass spectrometer to support drug metabolism and pharmacokinetic studies is described. The instrument is based on a quadrupole ion path and is capable of conventional tandem mass spectrometry (MS/MS) as well as several high-sensitivity ion trap MS scans using the final quadrupole as a linear ion trap. Several pharmaceutical compounds, including trocade, remikiren and tolcapone, were used to evaluate the capabilities of the system with positive and negative turbo ionspray, using either information-dependent data acquisition (IDA) or targeted analysis for the screening, identification and quantification of metabolites. Owing to the MS/MS in-space configuration, quadrupole-like CID spectra with ion trap sensitivity can be obtained without the classical low mass cutoff of 3D ion traps. The system also has MS(3) capability which allows fragmentation cascades to be followed. The combination of constant neutral loss or precursor ion scan with the enhanced product ion scan was found to be very selective for identifying metabolites at the picogram level in very complex matrices. Owing to the very high cycle time and, depending on the mass range, up to eight different MS experiments could be performed simultaneously without compromising chromatographic performance. Targeted product ion analysis was found to be complementary to IDA, in particular for very low concentrations. Comparable sensitivity was found in enhanced product ion scan and selected reaction monitoring modes. The instrument is particularly suitable for both qualitative and quantitative analysis.  相似文献   

7.
It can be argued that the last true paradigm shift in the bioanalytical (BA) arena was the shift from high-performance liquid chromatography (HPLC) with ultraviolet (UV) detection to HPLC with tandem mass spectrometry (MS/MS) detection after the commercialization of the triple quadrupole mass spectrometer in the 1990s. HPLC-MS/MS analysis based on selected reaction monitoring (SRM) has become the gold standard for BA assays and is used by all the major pharmaceutical companies for the quantitative analysis of new drug entities (NCEs) as part of the new drug discovery and development process. While LC-MS/MS continues to be the best tool for drug discovery bioanalysis, a new paradigm involving high-resolution mass spectrometry (HRMS) and ultrahigh-pressure liquid chromatography (uHPLC) is starting to make inroads into the pharmaceutical industry. The ability to collect full scan spectra, with excellent mass accuracy, mass resolution, 10-250 ms scan speeds and no NCE-related MS parameter optimization, makes the uHPLC-HRMS techniques suitable for quantitative analysis of NCEs while preserving maximum qualitative information about other drug-related and endogenous components such as metabolites, degradants, biomarkers and formulation materials. In this perspective article, we provide some insight into the evolution of the hybrid quadrupole-time-of-flight (Qq-TOF) mass spectrometer and propose some of the desirable specifications that such HRMS systems should have to be integrated into the drug discovery bioanalytical workflow for performing integrated qualitative and quantitative bioanalysis of drugs and related components.  相似文献   

8.
Triple quadrupole mass spectrometers are generally considered the instrument of choice for quantitative analysis. However, for the analysis of large peptides we have encountered some cases where, as the data presented here would indicate, ion trap mass spectrometers may be a good alternative. In general, specificity and sensitivity in bioanalytical liquid chromatography/mass spectrometry (LC/MS) assays are achieved via tandem MS (MS/MS) utilizing collision-induced dissociation (CID) while monitoring unique precursor to product ion transitions (i.e. selected reaction monitoring, SRM). Due to the difference in CID processes, triple quadrupoles and ion traps often generate significantly different fragmentation spectra of product ion species and intensities. The large peptidic analytes investigated here generated fewer fragments with higher relative abundance on the ion trap as compared to those generated on the triple quadrupole, resulting in lower limits of detection on the ion trap.  相似文献   

9.
A new type of quadrupole linear ion trap mass spectrometer, Q TRAP trade mark LC/MS/MS system (Q TRAP trade mark ), was evaluated for its performance in two studies: firstly, the in vitro metabolism of gemfibrozil in human liver microsomes, and, secondly, the quantification of propranolol in rat plasma. With the built-in information-dependent-acquisition (IDA) software, the instrument utilizes full scan MS in the ion trap mode and/or constant neutral loss scans as survey scans to trigger product ion scan (MS(2)) and MS(3) experiments to obtain structural information of drug metabolites 'on-the-fly'. Using this approach, five metabolites of gemfibrozil were detected in a single injection. This instrument combines some of the unique features of a triple quadrupole mass spectrometer, such as constant neutral loss scan, precursor ion scan and multiple reaction monitoring (MRM), together with the capability of a three-dimensional ion trap. Therefore, it becomes a powerful instrument for metabolite identification. The fast duty cycle in the ion trap mode allows the use of full product ion scan for quantification. For the quantification of propranolol, both MRM mode and full product ion scan in the ion trap mode were employed. Similar sensitivity, reproducibility and linearity values were established using these two approaches. The use of the product ion scan mode for quantification provided a convenient tool in selecting transitions for improving selectivity during the method development stage.  相似文献   

10.
The analytical suitabilities of three different liquid chromatography–tandem mass spectrometry (LC-MS/MS) systems, (1) triple quadrupole (QqQ), (2) conventional 3D ion trap (IT), and (3) quadrupole–linear IT (QqLIT), to determine trace levels of perfluorinated compounds (PFCs) in fish and shellfish were compared. Sample preparation was performed using alkaline extraction and solid-phase-extraction cleanup. This evaluation was focused on both quantitative (sensitivity, precision, and accuracy) and qualitative (identification capabilities) aspects. In the three instruments, the former facet was evaluated using selected reaction monitoring (SRM), which is the standard mode for quantitative LC-MS/MS analysis. Accuracy was similar in the three systems, with recoveries always over 70 %. Precision was better for the QqLIT and QqQ systems (7–15%) than for the IT system (10–17%). The QqLIT (working in SRM mode) and QqQ systems offered a linear dynamic range of at least 3 orders of magnitude, whereas that of the IT system was 2 orders of magnitude. The QqLIT system achieved at least 20-fold higher sensitivity than the QqQ system, and this was at least tenfold higher sensitivity than for the IT system. In the IT system, identification was based on sensitive full mass range acquisition and MS n fragmentation and in the QqLIT system, it was based on the use of an information-dependent-acquisition scan function, which allows the combination of an SRM or MS full scan acting as the survey scan and an enhanced product ion scan followed by MS3 as the dependent scan in the same analysis. Three instruments were applied to monitor the content in fish and shellfish (anchovies, swordfish, tuna, mussels, and oysters) obtained from Valencia and Barcelona markets (Spain). The eight target PFCs were detected at mean concentrations in the range from 10 ng ?kg-1 (perfluoro-7-methyloctanoic acid and perfluoro-1-decanesulfonate) to 4,200 ng ?kg-1 (perfluoropentanoic acid). Furthermore, perfluoroheptanoic and perfluoroundecanoic acids (not covered as target analytes) were identified in some samples.
Figure
PFCs in the environment  相似文献   

11.
The present contribution is focused on the evaluation of a high‐speed triple quadrupole mass spectrometer, carried out under moderately fast GC conditions (analysis time: 16.6 min). The mass spectrometric instrument can be operated under high‐speed GC conditions, in both full‐scan (maximum scan speed: 20 000 amu/s) and multiple reaction monitoring (MRM) modes (minimum dwell time: 0.01 s). Additionally, the mass spectrometric system can generate full scan and MRM information, simultaneously and rapidly. A headspace solid‐phase microextraction with fast GC coupled to triple quadrupole MS approach was developed for the: (i) qualitative untargeted analysis of brewed tea volatiles, and (ii) MRM qualitative and quantitative analysis of targeted volatiles (also in brewed tea), namely 30 phytosanitary contaminants. The performance of the triple quadrupole instrument was satisfactory both for identification and quantification purposes. Furthermore, the method sensitivity was more than sufficient for the requirements of current legislation. Method validation, related to the MRM analysis, was performed considering: precision of quantification data (maximum coefficient of variation value: 12.0%) and quantification/qualification ion ratios (maximum coefficient of variation value: 14.4%), along with limits of detection (4 parts per trillion–5 parts per billion range) and quantification (14 parts per trillion–16 parts per billion range).  相似文献   

12.
The use of the recently introduced Q-Trap mass spectrometer in the study of protein glycosylation is described. The combined ion trap and triple quadrupole scan functions make it a powerful system in both oligosaccharide and glycopeptide analysis. Several oligosaccharides, both linear and branched, were analyzed to obtain information on sequence, linkage, and branching. Quadrupole like MS/MS spectra with ion trap sensitivity but without the typical ion trap low mass cut-off were obtained. To determine the origin of fragments and to reveal the existence of new ions, the MS(3) capabilities of the system proved to be useful. Glycopeptides were selectively detected in peptide mixtures using the triple quadrupole precursor ion scan function, either in off-line experiments or during LC/MS using information dependent acquisition (IDA).  相似文献   

13.
Doubly protonated peptides that undergo an electron transfer reaction without dissociation in a linear ion trap can be subjected to beam-type collisional activation upon transfer from the linear ion trap into an adjacent mass analyzer, as demonstrated here with a hybrid triple quadrupole/linear ion trap system. The activation can be promoted by use of a DC offset difference between the ion trap used for reaction and the ion trap into which the products are injected of 12-16 V, which gives rise to energetic collisions between the transferred ions and the collision/bath gas employed in the linear ion trap used for ion/ion reactions. Such a process can be executed routinely on hybrid linear ion trap/triple quadrupole tandem mass spectrometers and is demonstrated here with several model peptides as well as a few dozen tryptic peptides. Collisional activation of the peptide precursor ions that survive electron transfer frequently provides structural information that is absent from the precursor ions that fragment spontaneously upon electron transfer. The degree to which additional structural information is obtained by collisional activation of the surviving singly charged peptide ions depends upon peptide size. Little or no additional structural information is obtained from small peptides (<8 residues) due to the high electron transfer dissociation (ETD) efficiencies noted for these peptides as well as the extensive sequence information that tends to be forthcoming from ETD of such species. Collisional activation of the surviving electron transfer products provided greatest benefit for peptides of 8-15 residues.  相似文献   

14.
15.
An ultra‐performance liquid chromatography/tandem mass spectrometry (UPLC/MS/MS) method for fast‐throughput analysis of eight anabolic and androgenic steroids (AAS) in equine plasma is reported. Analytes were recovered by liquid‐liquid extraction using methyl tert‐butyl ether, separated on a 1.9 µm C18 reversed‐phase column, and analyzed in positive electrospray ionization mode on a triple quadrupole mass spectrometer with selected reaction monitoring (SRM) and full product ion scans. Two SRM ion transitions were monitored for each AAS during screening to obtain highly selective screening results. Full product ion spectra of excellent quality for AAS, at 100 pg/0.5 mL in plasma, devoid of interfering spectra from impurities in plasma, were obtained. To our knowledge, this is the first report on the acquisition of full product ion spectra at such a low analyte concentration and plasma volume using a triple quadrupole instrument. In addition to product ion intensity ratios obtained from three SRM scans for identifying AAS in equine plasma, full product ion spectra were used as supporting evidence for confirmation. For quantification, deuterium‐labeled testosterone and stanozolol were used as internal standards (ISs). The limits of detection, quantification and confirmation were 6.25–12.5 pg/0.5 mL, 25 pg/0.5 mL and 50–100 pg/0.5 mL, respectively. There was no significant matrix effect on the analysis of all eight AAS. Intra‐day precision and accuracy were 2–15% and 91–107%, respectively. Inter‐day precision and accuracy were 1–21% and 94–110%, respectively. Total analysis time was 5 min. To date, the method has been successfully used in the analysis of >12 000 samples for AAS in plasma samples from racehorses competing in the State of Pennsylvania. The method is fast, selective, reproducible, and reliable. Copyright © 2009 John Wiley & Sons, Ltd.  相似文献   

16.
The purpose of the study was to determine the advantages of different mass spectrometric instruments and commercially available metabolite identification programs for metabolite profiling. Metabolism of tramadol hydrochloride and the excretion of it and its metabolites into human urine were used as a test case because the metabolism of tramadol is extensive and well known. Accurate mass measurements were carried out with a quadrupole time-of-flight mass spectrometer (Q-TOF) equipped with a LockSpray dual-electrospray ionization source. A triple quadrupole mass spectrometer (QqQ) was applied for full scan, product ion scan, precursor ion scan and neutral loss scan measurements and an ion trap instrument for full scan and product ion measurements. The performance of two metabolite identification programs was tested. The results showed that metabolite programs are time-saving tools but not yet capable of fully automated metabolite profiling. Detection of non-expected metabolites, especially at low concentrations in a complex matrix, is still almost impossible. With low-resolution instruments urine samples proved to be challenging even in a search for expected metabolites. Many false-positive hits were obtained with the automated searching and manual evaluation of the resulting data was required. False positives were avoided by using the higher mass accuracy Q-TOF. Automated programs were useful for constructing product ion methods, but the time-consuming interpretation of mass spectra was done manually. High-quality MS/MS spectra acquired on the QqQ instrument were used for confirmation of the tramadol metabolites. Although the ion trap instrument is of undisputable benefit in MS(n), the low mass cutoff of the ion trap made the identification of tramadol metabolites difficult. Some previously unreported metabolites of tramadol were found in the tramadol urine sample, and their identification was based solely on LC/MS and LC/MS/MS measurements.  相似文献   

17.
The use of a Q-q-Q(linear ion trap) instrument to obtain product ion spectra is described. The instrument is based on the ion path of a triple quadrupole mass spectrometer with Q3 operable as either a conventional RF/DC quadrupole mass filter or a linear ion trap mass spectrometer with axial ion ejection. This unique ion optical arrangement allows de-coupling of precursor ion isolation and fragmentation from the ion trap itself. The result is a high sensitivity tandem mass spectrometer with triple quadrupole fragmentation patterns and no inherent low mass cut-off. The use of the entrance RF-only section of the instrument as accumulation ion trap while the linear ion trap mass spectrometer is scanning enhances duty cycles and results in increased sensitivities by as much as a factor of 20. The instrument is also capable of all of the triple quadrupole scans including multiple-reaction monitoring (MRM) as well as precursor and constant neutral loss scanning. The high product ion scanning sensitivity allows the recording of useful product ion spectra near the MRM limit of quantitation.  相似文献   

18.
Transmission mode ion/ion reactions have been performed within the first quadrupole, the Q0 radiofrequency (RF)‐only quadrupole, of two types of hybrid tandem mass spectrometers (viz., triple quadrupole/linear ion trap and QqTOF instruments). These transmission mode reactions involved the storage of either the reagent species and the transmission of the analyte species through the Q0 quadrupole for charge inversion reactions or the storage of the analyte ions and transmission of the reagent ions as in charge reduction experiments. A key advantage to the use of transmission mode ion/ion reactions is that they do not require any instrument hardware modifications to provide interactions of oppositely charged ions and can be implemented in any instrument that contains a quadrupole or linear ion trap. The focus of this work was to investigate the potential of using the RF‐only quadrupole ion guide positioned prior to the first mass‐resolving element in a tandem mass spectrometer for ion/ion reactions. Two types of exemplary experiments have been demonstrated. One involved a charge inversion reaction and the other involved a charge reduction reaction in conjunction with ion parking. Ion/ion reactions proved to be readily implemented in Q0 thereby adding significantly greater experimental flexibility in the use of ion/ion reaction experiments with hybrid tandem mass spectrometers. Copyright © 2009 John Wiley & Sons, Ltd.  相似文献   

19.
目前,β-内酰胺类抗生素在临床抗感染药物中占有十分突出的地位,但在近年来的药品不良反应报告中,抗生素类药物引起的不良反应也占据了很高的比例,其中有我国生活环境影响、感染性疾病多的客观因素,但病人用药盲目性大、医生用药随意性多的问题也普遍存在。因此,进一步加强对抗生素类药物的监测,开发快速有效的分析测试方法显得十分重要。本文基于实际全血未知样品,开发了基于固相萃取及液质联用技术,快速准确地对血液中青霉素G及头孢拉定进行定性及质谱定量分析的检测方法。  相似文献   

20.
A triple-quadrupole instrument and a hybrid quadrupole/time-of-flight (TOF) mass spectrometer were compared for the determination of pharmaceutical compounds in water samples. The drugs investigated were the analgesics Ibuprofen, Fenoprofen, Ketoprofen, Naproxen, and Diclofenac. The recently introduced Q2-pulsing function, which enhances the transmission of fragment ions of a selected m/z window from the collision cell into the TOF analyzer, improved the sensitivity of product ion scans on the quadrupole/TOF instrument. The selectivity is much better on quadrupole/TOF systems than on triple quadrupoles because the high resolving power of the reflectron-TOF mass analyzer permits high-accuracy fragment ion selection. This minimizes interferences from environmental matrices and allows acquisition of full spectra for selected analytes with better signal-to-noise characteristics than comparable spectra obtained with a scanned quadrupole. The qualitative information obtained (mass accuracy, resolution and full-scan spectra) by hybrid quadrupole/TOF mass spectrometry allows a more certain identification of analytes in environmental matrices at trace levels. Sample enrichment of water samples was achieved by a solid-phase extraction procedure. Average recoveries for loading 1 L of samples varied from 88 to 110%, and the quantification limits were less than 1.2 ng/L for the triple-quadrupole instrument (in MRM mode) and less than 3 ng/L for the quadrupole/TOF instrument.  相似文献   