Overloading control of gas chromatographic systems
Authors:Igor G Zenkevich  Alexander A Pavlovskii
Affiliation:Institute for Chemistry, St. Petersburg State University, St. Petersburg, Russia
Abstract:Increasing the interlaboratory reproducibility of gas chromatographic retention indices requires avoiding measurements distorted by overloading effects. Several criteria of evaluating the limits of the mass overloading of gas chromatographic systems are compared and reconsidered. The criteria mostly appropriate for practical purposes are based on (i) the dependences of factors of peak broadening (ratio of peak height and its width) vs. amount of analyte injected into the chromatographic column and (ii) the dependence of parameters characterizing the peak distortion (asymmetry factor) vs. the amount of analyte. Both these criteria provide mutually comparable evaluations of the overloading limits for analytes of different polarity. At the same time, the dependence of retention indices vs. amounts of analyte injected in the chromatographic column cannot be recommended for overloading control, because the parameters of the corresponding linear regressions indicate temperature dependence. The interpretation of certain gas chromatographic anomalies requires the correct evaluation of overloading limits. For example, the unusual temperature dependence of retention indices of polar analytes on non‐polar stationary phases and the dependence of retention indices on ratio of amounts of target analytes and reference compounds.
Keywords:Gas chromatography  Overloading control  Overloading criteria