首页 | 官方网站   微博 | 高级检索  
     


Identification of isomers and control of ionization and dissociation processes using dual-mass-spectrometer scheme and genetic algorithm optimization
Authors:Chen Zhou;Tong Qiu-Nan;Zhang Cong-Cong;Hu Zhan
Affiliation:Chen Zhou;Tong Qiu-Nan;Zhang Cong-Cong;Hu Zhan;Institute of Atomic and Molecular Physics, Jilin University;
Abstract:Identification of acetone and its two isomers, and the control of their ionization and dissociation processes are performed using a dual-mass-spectrometer scheme. The scheme employs two sets of time of flight mass spectrometers to simultaneously acquire the mass spectra of two different molecules under the irradiation of identically shaped femtosecond laser pulses. The optimal laser pulses are found using closed-loop learning method based on a genetic algorithm. Compared with the mass spectra of the two isomers that are obtained with the transform limited pulse, those obtained under the irradiation of the optimal laser pulse show large differences and the various reaction pathways of the two molecules are selectively controlled. The experimental results demonstrate that the scheme is quite effective and useful in studies of two molecules having common mass peaks, which makes a traditional single mass spectrometer unfeasible.
Keywords:
本文献已被 CNKI 等数据库收录!