首页 | 官方网站   微博 | 高级检索  
     


Simultaneous determination of plasma creatinine,uric acid,kynurenine and tryptophan by high‐performance liquid chromatography: method validation and in application to the assessment of renal function
Authors:Jianxing Zhao
Affiliation:Ruijin Hospital, Shanghai Jiao Tong University School of Medicine and Shanghai Institute of Hypertension, Shanghai, China
Abstract:A high‐performance liquid chromatography with ultraviolet detection method has been developed for the simultaneous determination of a set of reliable markers of renal function, including creatinine, uric acid, kynurenine and tryptophan in plasma. Separation was achieved by an Agilent HC‐C18 (2) analytical column. Gradient elution and programmed wavelength detection allowed the method to be used to analyze these compounds by just one injection. The total run time was 25 min with all peaks of interest being eluted within 13 min. Good linear responses were found with correlation coefficient >0.999 for all analytes within the concentration range of the relevant levels. The recovery was: creatinine, 101 ± 1%; uric acid, 94.9 ± 3.7%; kynurenine, 100 ± 2%; and tryptophan, 92.6 ± 2.9%. Coefficients of variation within‐run and between‐run of all analytes were ≤2.4%. The limit of detection of the method was: creatinine, 0.1 µmol/L; uric acid, 0.05 µmol/L; kynurenine, 0.02 µmol/L; and tryptophan, 1 µmol/L. The developed method could be employed as a useful tool for the detection of chronic kidney disease, even at an early stage. Copyright © 2014 John Wiley & Sons, Ltd.
Keywords:creatinine  uric acid  kynurenine  tryptophan  plasma  high‐performance liquid chromatography
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号