首页 | 本学科首页   官方微博 | 高级检索  
     


Scattering correction method for panel detector based cone beam computed tomography system
Authors:Jia Peng-Xiang  Zhang Feng  Yan Bin  Bao Shang-Lian
Affiliation:Beijing City Key Lab of Medical Physics and Engineering, Peking University, Beijing 100871, China; China National Digital Switching System Engineering and Technological Research Center, Zhengzhou 450002, China
Abstract:A scattering correction method for a panel detector based cone beam computed tomography system is presented. First, the x-ray spectrum of the system is acquired by using the Monte Carlo simulation method. Secondly, scattered photon distribution is calculated and stored as correction matrixes by using the Monte Carlo simulation method according to scanned objects and computed tomography system specialties. Thirdly, scattered photons are removed from projection data by correction matrixes. A comparison of reconstruction image between before and after scattering correction demonstrates that the scattering correction method is effective for the panel detector based cone beam computed tomography system.
Keywords:scattering correction  Monte Carlo simulation  panel detector  cone beam computed tomography system
本文献已被 CNKI 维普 等数据库收录!
点击此处可从《中国物理 B》浏览原始摘要信息
点击此处可从《中国物理 B》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号